Why does my blender leak gas out the bottom?

Can a blender be specially configured to meet my requirements?

Do I need a Service Requisition Form to send a unit in for service?


 

Why does my blender leak gas out the bottom?

The blender only exhibits this bleed when an auxiliary port, labeled as such on the side, is used. This bleed is necessary to ensure concentration accuracy when using the blender for low flows, typically between 3 and 15 lpm. If your flow requirements are above 15 lpm, then the primary port is recommended as this port does not induce a bleed.
If your flow requirements are always less than 30 lpm, the 2003 Low Flow Blender is recommended. This blender has no bleed from 3 to 30 lpm and for flows less than 3 lpm, only a 3 lpm bleed is induced. Click here for more on this subject.



Back

Can a blender be specially configured to meet my requirements?

Yes. Special fittings can be adapted as well as custom face plates using your own artwork. Special mounting needs can also be addressed. For more information, contact Bio-Med Devices directly.



Back

Do I need a Service Requisition Form  to send a unit in for service?

Yes.  Download Service Requisition Form .